The uniTOC device uses the full oxidation by UV. The oxidation process can be enhanced with additional dosage of reagent (persulfate). By using the NDIR detector the uniTOC gives precise results in applications of ground water to ultrapure water. The NDIR detector is absolutely selective to carbon dioxide and therefore independent from conductivity and sample temperature. Measuring times are short and the continuous self-cleaning of the reactor avoids any cross-contamination.
Since the quality of the stripped gas can be low, the use of pressurised air in combination with our air filters saves money and is optionally available.
- METHOD & DETECTION
SAMPLE VOLUME FLOW
-TOTAL ORGANIC CARBON (TOC)-RANGE
-RELATED NORMS
1 | Method & detection | UV oxidation & NDIR (enhanced oxidation; configurable with reagent) | |
---|---|---|---|
2 | Sample volume | 20 mL loop (low range), 5 mL loop (high range) | |
3 | Period of measurement | 6 to 15 min | |
4 | Measurement range | 2 to 5000 ppb (low range), 15 ppb to 20 ppm (high range) | |
5 | Detection accuracy | ~3 ppb or 2 % (whatever is greater) | |
6 | Ports | 2 for samples | 1 for reagent | |
7 | Related norms | USP <643>, DIN 32645, EP 2.2.44, DIN EN 1484, CFR 21 part 11, IP 45 | |
8 | Ambient parameters | Relative humidity: 10 to 95 %, temperature: 10 to 45 °C | |
9 | Interface ports | analog (4 - 20 mA) | |
10 | Dimensions, weight, power | 400 x 300 x 600 mm, 24 kg, 110 - 230 V, 50 Hz, 70 to 105 Watt |
1 | Type of sample medium | From ground water to ultra pure water |
---|---|---|
2 | Sample temperature | 10 to 50 °C |
3 | Inlet pressure | < 1.5 bar |
4 | Compressed gas | air, oxygen, nitrogen | free of carbon dioxide |
5 | Inlet pressure of gas | Nitrogen: 1 to 1.5 bar | oxygen: 3.5 to 5 bar |