UniTOC

The uniTOC device uses the full oxidation by UV. The oxidation process can be enhanced with additional dosage of reagent (persulfate). By using the NDIR detector the uniTOC gives precise results in applications of ground water to ultrapure water. The NDIR detector is absolutely selective to carbon dioxide and therefore independent from conductivity and sample temperature. Measuring times are short and the continuous self-cleaning of the reactor avoids any cross-contamination.

Since the quality of the stripped gas can be low, the use of pressurised air in combination with our air filters saves money and is optionally available.

Technical Specifications -

  •  UV oxidation (full) & NDIR
  • - METHOD & DETECTION

  •  20 mL (low range), 5 mL loop (high range) -
  • SAMPLE VOLUME FLOW

  •  2 to 5000 ppb (low range), 15 ppb to 20 ppm (high range)-
  • -TOTAL ORGANIC CARBON (TOC)-RANGE

  •  USP <643>, EP 2.2.44, DIN EN 1484, DIN 32645, 21 CFR part 11, IP 45-
  • -RELATED NORMS

UniTOC Advantages -

  •  NDIR DETECTOR / TOC IS DETECTED DIRECTLY OVER CO2 AND NOT VIA A 3RD PARAMETER (CONDUCTIVITY). THEREFORE NO LIMIT TO CONDUCTIVITY.
  •   9 MONTHS UV LAMP LIFE GUARANTEE / 50% LONGER LIFE TIME RANGE THAN USUAL.
  •  AVAILABLE FOR TWO RANGES OF TOC.
  •  2 SAMPLE INLETS / 2 POINTS OF THE LOOP CAN BE MEASURED ,SO DOUBLE EFFICIENCY.
  •  REMOTE SERVICE / INSTRUMENT CAN BE CONNECTED TO INTERNET TO GET SERVICE DIAGNOSIS

Options

Specifications -

Technical Specifications -
1 Method & detection UV oxidation & NDIR (enhanced oxidation; configurable with reagent)
2 Sample volume 20 mL loop (low range), 5 mL loop (high range)
3 Period of measurement 6 to 15 min
4 Measurement range 2 to 5000 ppb (low range), 15 ppb to 20 ppm (high range)
5 Detection accuracy ~3 ppb or 2 % (whatever is greater)
6 Ports 2 for samples | 1 for reagent
7 Related norms USP <643>, DIN 32645, EP 2.2.44, DIN EN 1484, CFR 21 part 11, IP 45
8 Ambient parameters Relative humidity: 10 to 95 %, temperature: 10 to 45 °C
9 Interface ports analog (4 - 20 mA)
10 Dimensions, weight, power 400 x 300 x 600 mm, 24 kg, 110 - 230 V, 50 Hz, 70 to 105 Watt
Medium Requirements -
1 Type of sample medium From ground water to ultra pure water
2 Sample temperature 10 to 50 °C
3 Inlet pressure < 1.5 bar
4 Compressed gas air, oxygen, nitrogen | free of carbon dioxide
5 Inlet pressure of gas Nitrogen: 1 to 1.5 bar | oxygen: 3.5 to 5 bar