Aquinity² models produce ultra-pure water directly from tap water. The Aquinity² P10 ultrapure water system is equipped with reverse osmosis (RO) module to produce deionized water directly into an integrated 10l tank. In the next step Type, I grade water is produced with a polisher cartridge. The Aquinity² water purification system is available in different configurations (Reagent & Analytical) to meet the specific requirements of ultra-pure water for different applications.
Analytical is a version with a UV lamp for low TOC value. Life Science grade water with low endotoxin can be generated with a special final filter. An optional integrated TOC sensor allows the user to monitor the TOC content permanently online. Besides that, the Aquinity² can be configured as a bench top or bench-integrated system, where the control panel will mount on the wall.
- RESISTIVITY/CONDUCTIVITY
PRODUCTION RATE
-TOTAL ORGANIC CARBON (TOC)
1 | ltrapure water quality | 0.055 µS/cm; Type I |
---|---|---|
2 | Resistivity | 18.2 MegaOhm/cm |
3 | Total organic carbon (TOC) | < 10 ppb reagent, < 3 ppb Analytical |
4 | Dispensing flow rate | 1.5 L/min |
5 | Production rate | 8 L/min* |
6 | Bacteria | < 1 cfu/mL |
7 | Particulate | > 0.2 µm less than 1 particulate/mL |
8 | RNAse | < 1 pg/mL** |
9 | DNAse | < 5 pg/mL** |
10 | Dimensions, weight, power | 504 x 340 x 535 mm, 16-20 kg, 110 - 230 V *not possible to tap pure water **with microbiological final filter |
1 | Type of feed water | Tap water |
---|---|---|
2 | Feed water conductivity | < 1400 µS/cm |
3 | Inlet pressure | 1.5 to 6 bar |
4 | Free chlorine | < 0.1 mg/L |
5 | Silt Density Inded (SDI) | < 3 |
6 | pH | 3 to 9 |
7 | Temperature | 5 to 25 °C |