miniTOC

Short Overview

The miniTOC instrument uses the classical flow-through concept for online or offline TOC monitoring of pure and ultra pure water.

The efficient UV oxidation of organic compounds by the “direct surface” principle in combination with conductivity detection guarantees the fast monitoring of values.

Technical Specifications:

UV-Oxidation (partial) & conductivity

Method & detection

ca. 14
mL/min

Sample volume flow

0,5 to 1000
ppb

Total Organic Carbon (TOC)-Range

USP <643>, EP 2.2.44, DIN EN 1484

Related norms

CE
Certified

Certified

miniTOC Advantages

Low sample volume / lower cost and waste

9 months UV lamp life guarantee / 50 % longer life time range than usual

Continuous measurement (every 2 seconds) / fast reaction

Offline measurement / grab samples can be measured

Up to 3 sample inlets / very efficient

Remote service / Instrument can be connected to internet to get service diagnosis)

Morning Dew
Morning Dew

Calibration & System Suitability Test (SST)

Calibration and SST of the miniTOC is easy and can be done by the user. Online operation is interrupted briefly for the system checks to be performed. The operator simply has to connect the inlet with the calibration standard solutions and the diluting water. The software enables the user to perform calibration and SST independently and the programmable sequence ensures the automatic procedure and data evaluation of the test. The verification of the TOC results according to the requirements of USP and Ph.Eur is realised by the periodic performance of system suitability tests.

Options

intervall mode

The miniTOC can be ordered as configuration with 4 inputs (miniTOC pro). With an additional software upgrade option the user can measure up to 3 loops sequentially.

Offline Version & autosampler

An offline configuration is available for the miniTOC. By using the offline mode of the miniTOC pro the system is also upgradable with an autosampler with 15 positions for 100 mL vials.

Check of sample flow

nitiates alarm in case of inlet flow is zero.

Heat exchanger

If the medium temperature is higher than 50 °C a heat exchanger is required.

Pressure regulator

If the medium pressure is higher than 1 bar a pressure transducer is required.

21 CFR part 11 Software

Software with 21 CFR part11 conformity for pharma applications.

Specifications

Technical Specifications
1 Method & detection UV oxidation (partial) & conductivity
2 Sample volume flow Approx. 14 mL/min
3 Time of measurement 2 sec (online mode) | 1 min (offline mode)
4 Measurement range 0.5 to 1000 ppb
5 Detection accuracy ~0.5 ppb or 1 % (whatever is greater)
6 Ports 1 for sample, 1 for SST (basic) | 1 for sample, 3 for SST (pro)
7 Related norms USP <643>, EP 2.2.44, DIN EN 1484
8 Ambient parameters Relative humidity: 10 to 95 %, temperature: 10 to 45 °C
9 Interface ports Analog, according to Namur NE43
9 Dimensions, weight, power 300 x 200 x 500 mm, 14 kg, 110 - 230 V, 50 Hz, 60 Watt
Technical Specifications
1 Type of sample medium Pure or ultra pure water
2 Medium conductivity < 2 µS/cm (upgradable to < 30 µS/cm)
3 Inlet pressure < 1 bar
4 Sample temperature < 50 °C